The US Food and Drug Administration (FDA) has designated the recall of Abbott’s CentriMag Acute Circulatory Support System as Class I or the most serious type.

Abbott’s device pumps blood through a patient for up to six hours during open-heart procedures. Furthermore, the CentriMag System offers temporary blood circulatory support for up to 30 days for patients in cardiogenic shock.

Abbott began a recall of the device in August over a calibration system error caused by electromagnetic interference that could cause the pump to slow or stop.

The interference may also cause the console screen to blank and result in inaccurate alarms.

The FDA added that slowing or stopping of the pump is associated with the risk of health complications including stroke, severe organ damage and even death.

When Abbott issued the recall in August, 44 injuries and one death due to the CentriMag System were reported. Of the total devices involved in the recall, 381 were in the US and 283 in other countries.

How well do you really know your competitors?

Access the most comprehensive Company Profiles on the market, powered by GlobalData. Save hours of research. Gain competitive edge.

Company Profile – free sample

Thank you!

Your download email will arrive shortly

Not ready to buy yet? Download a free sample

We are confident about the unique quality of our Company Profiles. However, we want you to make the most beneficial decision for your business, so we offer a free sample that you can download by submitting the below form

By GlobalData
Visit our Privacy Policy for more information about our services, how we may use, process and share your personal data, including information of your rights in respect of your personal data and how you can unsubscribe from future marketing communications. Our services are intended for corporate subscribers and you warrant that the email address submitted is your corporate email address.

In its recall letter, the company recommended the use of the motor until recalibration can be performed if it is not showing electromagnetic interference issues.

In the case of electromagnetic interference​, notified by alarm, the pump should be switched to the backup system, the company added.

Other recommendations were moving the system or console away from the electromagnetic interference sources and removing the source of the interference.

Abbott has since made changes to the motor calibration process in manufacturing to alleviate the problem.